Martel MC-1210 Multifunction Calibrator

Harga :
CALL
Merk :
Martel
Pengiriman :
3 hari sejak pembayaran
Lokasi :
Dilihat :
191 Kali
Update Terakhir :
01-12-2014 17:25

Detail Produk

 The MC-1210 is a rugged and reliable universal multifunction calibrator. It’s dual display and isolated readback circuit allows it to power a transmitter under test while reading its milliamp output. Truly an all-in-one calibrator. The MC-1210 Multifunction Calibrator also has a wide range of switch test features for both pressure and temperature switches.
  • Dual Display. The upper display is used for the measurement of volts, current, and pressure. The lower display can be used to measure volts, current, pressure, resistance temperature detectors (RTD’s), thermcouples (TC’s), frequency, resistance, and to source pulse trains.
  • A thermocouple (TC) input/output terminal with automatic reference-junction temperature compensation
  • Setpoint in each range for quickly increasing/decreasing output
  • Complete serial interface for remote control
  • Isolated read back for transmitter calibrations
  • Switchtest capability
Features:
  • Multifunction
  • Dual Display
  • Isolated Readback Circuit
  • ClearbriteTM Graphic Display
  • Auto Stepping
  • Auto Ramping
  • RS-232 Serial Interface
  • NIST Calibration Certificate
  • Rubber Boot
  • Martel 3-Key User Interface
  • Scroll/Step Output
  • Numeric Input
  • AC Charger/Adapter Option
Specifications (All specifications apply at 23 °C ± 5°C unless otherwise stated. Outside this range add ±0.005% of reading/°C. Accuracy is % of reading ± floor spec.)

General Specifications (10 Series)

   
Operating Temperature -10 °C to 50 °C
Storage Temperature -20 °C to 70 °C
Power 4 AA Alkaline or optional rechargeable batteries
Low Battery Warning Yes, on display
ClearBrite™ Display High contrast 128 x 64 pixel addressable graphic LCD w backlight
2.4" x 1.8" (63 mm x 44 mm)
Serial Communications Yes, ASCII, RS-232, requires optional 1919069 serial cable or 1919896 USB cable
CE-EMC EN50082-1 and EN55022: 1994 Class B
Safety CSA C22.2 No. 1010-1: 1992
Weight 1.8 lb (0.82 kg)
Size 8.5" x 4.8" x 2.1" (22 cm x 12 cm x 5.3 cm)
Other IP54 
Protected against misconnection to 250 VAC/VDC

DC Voltage & Current Measure/Source

Note: optional 50 mA range available          
  DC Voltage Upper Isolated Measurement DC Voltage Lower Non-Isolated Measure DC Voltage Lower Non-Isolated Source DC Current Upper Isolated DC Current Lower Isolated Measurement DC Current Lower Isolated Source
Range 0.000V - 30.000V 0.000V - 20.000V 0.000V - 20.000V 0.000 mA - 24.000 mA 0.000 mA - 24.000 mA 0.000 mA - 24.000 mA
Accuracy 0.015% ± 2 mV 0.015% ± 2 mV 0.015% ± 2 mV 0.015% ± 2 uA 0.015% ± 2 uA 0.015% ± 2 uA

Frequency Measure/Source

             
  Frequency Measurement Frequency Source Frequency Measurement Frequency Source Frequency Measurement Frequency Source
Range 2.0 CPM - 600.0 CPM 2.0 CPM - 600.0 CPM 1.0 Hz - 1000.0 Hz 1.0 Hz - 1000.0 Hz 1.00 kHz - 10.00 kHz 1.00 kHz - 10.00 kHz
Accuracy 0.05% ± 0.1 CPM 0.05% 0.05% ± 0.1 Hz 0.05% 0.05% ± 0.01 kHz 0.05%

Series Resistance Measure (Autoranging)

     
  Resistance Ohms Low Resistance Ohms High
Range 0.00Ω - 400.00Ω 401.0Ω - 4000.0Ω
Accuracy 0.025% ± 0.05Ω 0.025% ± 0.5Ω

Series Resistance Source (Autoranging)

         
Resistance Source Ohms Source Low Excitation Current
0.1 -0.5 mA
Ohms Source Low Excitation Current
0.5 - 3 mA
Ohms Source High Excitation Current
0.05 - 0.8 mA
Ohms Source High Excitation Current
0.05 - 0.4 mA
Range 5.0Ω - 400.00Ω 5.0Ω - 400.00Ω 401.0Ω - 1500Ω 1500Ω - 4000Ω
Accuracy 0.025% ± 0.1Ω 0.025% ± 0.05Ω 0.025% ± 0.5Ω 0.025% ± 0.5Ω

RTD Measure/Source

Accuracy statements are based on 4W connections    
RTD Type Range (°C) Accuracy (°C)
PT385, 10Ω -200.0 - -80.0
-80.0 - 0.0
0.0 - 100.0
100.0 - 300.0
300.0 - 400.0
400.0 - 630.0 
630.0 - 800.0
1.3
1.3
1.4
1.5
1.6
1.8
1.9
PT385, 50Ω -200.0 - -80.0
-80.0 - 0.0
0.0 - 100.0
100.0 - 300.0
300.0 - 400.0
400.0 - 630.0 
630.0 - 800.0
0.3
0.4
0.4
0.4
0.5
0.5
0.6
PT385, 100Ω -200.0 - -80.0
-80.0 - 0.0
0.0 - 100.0
100.0 - 300.0
300.0 - 400.0
400.0 - 630.0 
630.0 - 800.0
0.1
0.2
0.2
0.2
0.3
0.3
0.4
PT3926, 100Ω -200.0 - -80.0
-80.0 - 0.0
0.0 - 100.0
100.0 - 300.0
300.0 - 400.0
400.0 - 630.0
0.1
0.2
0.2
0.2
0.3
0.3
PT3916, 100Ω -200.0 - -80.0
-80.0 - 0.0
0.0 - 100.0
100.0 - 260.0
260.0 - 300.0
300.0 - 400.0 
400.0 - 630.0
0.1
0.2
0.2
0.2
0.3
0.3
0.3
PT385, 200Ω -200.0 - -80.0
-80.0 - 0.0
0.0 - 100.0
100.0 - 300.0
300.0 - 400.0
400.0 - 630.0
0.6
0.7
0.7
0.7
0.7
0.9
0.9
PT385, 500Ω -200.0 - -80.0
-80.0 - 0.0
0.0 - 100.0
100.0 - 300.0
300.0 - 400.0
400.0 - 630.0
0.2
0.3
0.3
0.4
0.4
0.4
0.5
PT385, 1000Ω -200.0 - -80.0
-80.0 - 0.0
0.0 - 100.0
100.0 - 300.0
300.0 - 400.0
400.0 - 630.0
0.2
0.2
0.2
0.2
0.3
0.3
0.4
NI120 -80.0 - 260.0 0.1
Cu10 -100.0 - 260.0 1.3
Cu50 -180.0 - 200.0 0.3
CU100 -180.0 - 200.0 0.1
YSI400 15.0 - 50.0 0.1

MilliVolts Measure/Source

Maximum current output in voltage ranges is 1mA with an output impedance of > 1Ω    
  Millivolts Measure Millivolts Source
Range -10.000 mV - 75.000 mV -10.000 mV - 75.000 mV
Accuracy 0.02% ± 10 uV 0.02% ± 10 uV

TC Measure/Source

       
TC Type Range (°C) Accuracy (°C)
with CJC OFF
Accuracy (°C)
with CJC ON
J -210.0 - 0.0
0.0 - 800.0
800.0 -1200.0
0.4
0.2
0.3
0.6
0.4
0.5
K -200.0 - 0.0
0.0 - 1000.0
1000.0 - 1372.0
0.6
0.3
0.5
0.8
0.5
0.7
T -250.0 - 0.0
0.0 - 400.0
0.6
0.2
0.8
0.4
E -250.0 - -100.0
-100.0 - 1000.0
0.6
0.2
0.8
0.4
R 0 - 1767 1.2 1.4
S 0 - 1767 1.2 1.4
B 600 - 800
800 - 1000
1000 - 1820
1.2
1.3
1.6
1
1.5
1.7
C 0 - 1000
1000 - 2316
0.6
2.3
0.8
2.5
XK -200.0 - 800.0 0.2 0.4
BP 0.0 - 800.0
800.0 - 2500.0
0.9
2.3
1.1
2.5
L -200.0 - 0.0
0.0 - 900.0
0.3
0.2
0.5
0.4
U -200.0 - 0.0 0.3 0.7
N 0.0 - 600.0
-200.0 - 0.0
0.0 -1300.0
0.3
0.8
0.4
0.5
1.0
0.6

Permintaan Penawaran