- More precise measurement and calibration source performance, accuracies of 0.01%.
- Two separate channels; measure, source and view process signals simultaneously
- Measure volts, mA, RTDs, thermocouples, frequency, and resistance to test sensors and transmitters
- Source/simulate volts, mA, thermocouples, RTDs, frequency, and pressure to calibrate transmitters
- Measure or source pressure* using any of 29 Fluke 700Pxx Pressure Modules
- Source mA with simultaneous pressure measurement to conduct valve and I/P tests
- Integrated pressure switch test allows you to capture the set, reset and deadband of a switch
- Memory storage for up to 8 calibration results, return stored calibration data from the field for later analysis
- Transmitter error % calculation, interpret calibration results without a calculator
- Perform fast linearity tests with auto step and auto ramp features
- Power transmitters during test using 24 V loop supply and simultaneous mA measurement
- Store frequently-used test setups for later use
- Frequency totalizer and frequency pulse train source mode for enhanced flowmeter testing
- HART mode inserts 250 ohm resistor in mA measure and source for compatibility with HART instrumentation
- Custom RTD curves, add calibration constants for certified RTD probes for enhanced temperature measurement.
- New voltage input protection design for improved reliability
- Three-year warranty
Measurement Accuracy | |||
Voltage DC | 30.000 V | 0.01% + 2 mV (upper display) | |
20.000 V | 0.01% + 2 mV (lower display) | ||
90.00 mV | 0.01% + 10 μV | ||
-10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | ||
Current DC | 24.000 mA | 0.01%+ 2 mA | |
Resistance | 0.0 to 4000.0 Ω | ± 0.015% | |
Frequency | 2.0 to 1000.0 CPM | 0.05% + 0.1 cpm | |
1.0 to 1000.0 Hz | 0.05% + 0.1 Hz | ||
1.00 to 15.00 kHz | 0.05% + 0.01 kHz | ||
Sensitivity | 1 V peak-to-peak-minimum | ||
Pressure |
Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure. |
||
Source Accuracy | |||
Voltage DC | 100.00 mV | 0.01% + 10 μV | |
20.000 V | 0.01% + 2 mV | ||
-10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | ||
Current DC | 24.000 mA (Source) | 0.01% + 2 μA | |
24.000 mA (Simulate) | 0.01% + 2 μA | ||
Resistance | Range, accuracy (percent of reading + floor) and excitation current | 5 to 400 Ω, 0.015% + 0.1 Ω, 0.1 to 0.5 mA | |
5 to 400 Ω, 0.015% + 0.05 Ω, 0.5 to 3 mA | |||
400 to 1500 Ω, 0.015% + 0.5 Ω, 0.05 to 0.8 mA | |||
1500 to 4000 Ω, 0.015% + 0.5 Ω, 0.05 to 0.4 mA | |||
Frequency | 2.0 to 1000.0 CPM | 0.05% | |
1.0 to 1000.0 Hz | 0.05% | ||
1.0 to 10.0 kHz | 0.25% | ||
10.0 to 15.0 kHz | 0.5% | ||
Waveform | 5 V p-p squarewave, -0.1 V offset | ||
RTDs and Thermocouples | |||
Measure and source accuracy | Cu10 | 1.8°C | |
NI-120 | 0.15°C | ||
PT-100 (385) | 0.45°C | ||
PT-100 (392) | 0.35°C | ||
PT-100 (391-JIS) | 0.35°C | ||
PT-200 (385) | 0.95°C | ||
PT-500 (385) | 0.55°C | ||
PT-1000 (385) | 0.35°C | ||
RTD Resolution | 0.01°C (0.01°F) | ||
Cu10.01 C | 0.1°C (0.1°F) | ||
J | 0.4°C | ||
K | 0.5°C | ||
T | 0.4°C | ||
E | 0.4°C | ||
R | 1.4°C | ||
S | 1.4°C | ||
B | 1.4°C | ||
L | 0.45°C | ||
U | 0.45°C | ||
N | 0.6°C | ||
XK | 0.4°C | ||
BP | 1.1°C | ||
TC Resolution | J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F | ||
Specifications | |||
Ramp functions | Source functions | Voltage, current, resistance, frequency, temperature | |
Ramps | Slow ramp, Fast ramp, 25% step-ramp | ||
Loop power function | Voltage | 24 V | |
Accuracy | 10% | ||
Maximum current | 22 mA, short circuit protected | ||
Step functions | Source functions | Voltage, current, resistance, frequency, temperature | |
Steps | 25% of range, 100% of range | ||
Environmental Specifications | |||
Operating temperature | -10°C to +55°C | ||
Storage temperature | -20°C to +71°C | ||
Operating altitude | 3000 m | ||
Safety Specifications | |||
Agency approvals | EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 | ||
Mechanical and General Specifications | |||
Size | 200 x 96 x 47 mm (7.9 x 3.8 x 1.9 in) | ||
Weight | 650 g (23 oz) | ||
Batteries | 4 AA alkaline batteries | ||
Battery replacement | Separate battery compartment, accessible without breaking calibration seal | ||
Side port connections | Pressure module connector, also used for remote real-time programming |