Fluke 741B Documenting Process Calibrator

Harga :
CALL
Merk :
Fluke
Pengiriman :
3 hari sejak pembayaran
Lokasi :
Dilihat :
183 Kali
Update Terakhir :
01-12-2014 16:20

Detail Produk

 
The 741B is easy to use with flexible test strategies. And like all Fluke tools, it's rugged, reliable and tough as nails. It's like having several powerful tools in one - so you can work faster than ever:
  • Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments
  • Source/simulate volts, mA, thermocouples , RTDs, frequency, ohms, and pressure to calibrate transmitters
  • Power transmitters during test using loop supply with simultaneous mA measurement
  • Measure/source pressure using any of 29 Fluke 700Pxx Pressure Modules
  • Measure and source simultaneously with one compact, rugged, reliable tool
  • Create and run automated as-found/as-left procedures to satisfy quality programs or regulations. Record and document results
  • Use many features like autostep, custom units, user entered values during test, one-point and two-point switch testing, square root DP flow testing, programmable measurement delay etc.
  • Easy to use
  • Three year warranty
  • Bright white display. Read both sourced and measured parameters.
  • Multi lingual interface
  • Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 1msec.
Measurement Accuracy
Voltage DC
110.000 mV  0.025%+0.015% *
1.10000 V  0.025%+0.005% *
11.0000 V  0.025%+0.005% *
110.000 V  0.05%+0.005% *
300.00 V  0.05%+0.005% *
Voltage AC
20 to 40 Hz  2% + 10
40 to 500 Hz  0.5% + 5
500 to 1 kHz  2% + 10
1 kHz to 5 kHz  10% + 20
Ranges:  1.1000, 11.000, 110.00, 300V
Current DC
30.000 mA  0.01% + 0.015% *
110.00 mA  0.01% + 0.015% *
Resistance
11.000 Ω  0.05% + 50 mΩ
110.00 Ω  0.05% + 50 mΩ
1.1000 kΩ  0.05% + 500 mΩ
11.000 kΩ  0.1% + 10 Ω
Frequency
1.00 to 109.99 Hz  0.05 Hz
110.0 to 1099.9 Hz  0.5 Hz
1.100 to 10.999 kHz  5 Hz
11.00 to 50.00 kHz  50 Hz
Pressure
Accuracy  from 0.025% of range using any of 29 pressure modules.
   (for detailed specifications refer to pressure modules in options and accessories)
   Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
Note
*  (% of reading + % of full scale)

 

Source Accuracy
Voltage DC
110.000 mV  0.01%+0.005% *
1.10000 V  0.01%+0.005% *
15.0000 V  0.01%+0.005% *
Current DC
22.000 mA (Source)  0.01%+ 0.015% *
22.000 mA (Simulate)  0.02% + 0.03% *
Resistance
11.000 Ω  0.01% + 20 mΩ
110.00 Ω  0.01% + 40 mΩ
1.1000 kΩ  0.02% + 500 mΩ
11.000 kΩ  0.03% + 5Ω
Frequency
0.00 to 10.99 Hz  0.01 Hz
11.00 to 109.99 Hz  0.1 Hz
110.0 to 1099.9 Hz  0.1 Hz
1100 to 21999 Hz  2 Hz
22.000 to 50.000 kHz  5 Hz
Note
*  (% of reading + % of full scale)

 

RTDs and Thermocouples
Measure Accuracy:
10 Ω Cu (427):  2 °C
100 Ω Pt (3916):  0.3 °C
100 Ω Pt (3926):  0.3 °C
100 Ω Pt (385):  0.3 °C
200 Ω Pt (385):  0.3 °C
500 Ω Pt (385):  0.3 °C
1000 Ω Pt (385):  0.3 °C
120 Ω Ni (672):  0.3 °C
note:  For 2-and 3-wire measurement, add 0.4 °C
E:  0.3 °C
N:  0.5 °C
J:  0.3 °C
L:  0.3 °C
K:  0.3 °C
T:  0.3 °C
U:  0.3 °C
B:  0.9 °C
R:  1.0 °C
S:  0.9 °C
C:  0.6 °C
note:  Accuracy with external cold junction, for internal junction add 0.2°C
Source Accuracy:
10 Ω Cu (427):  1 °C
100 Ω Pt (3916):  0.1 °C
100 Ω Pt (3926):  0.1 °C
100 Ω Pt (385):  0.1 °C
200 Ω Pt (385):  0.1 °C
500 Ω Pt (385):  0.1 °C
1000 Ω Pt (385):  0.1 °C
120 Ω Ni (672):  0.1 °C
note:  For 2-and 3-wire simulation, add 0.4 °C
E:  0.2 °C
N:  0.3 °C
J:  0.2 °C
L:  0.2 °C
K:  0.3 °C
T:  0.3 °C
U:  0.3 °C
B:  0.8 °C
R:  0.9 °C
S:  0.9 °C
C:  0.6 °C
note:  Accuracy with external cold junction, for internal junction add 0.2 °C

 

Specifications
Ramp Functions
Source functions:  Voltage, current, resistance, frequency, temperature
Rate:  4 steps/second
Trip detect:  Continuity or voltage (continuity detection not available when sourcing current)
Loop Power Function
Voltage:  Selectable, 24 V or 28 V
Accuracy:  5%
Maximum current:  22 mA, short circuit protected
Maximum input voltage:  30 V DC
Step Functions
Source Functions  Voltage, current, resistance, frequency, temperature
Manual Step  Selectable step, change with arrow buttons
Autostep  Fully programmable for function, start delay, stepvalue, time per step, repeat

 

Environmental Specifications
Operating Temperature
   -10 °C to +50 °C
   -20 °C to +50 °C *
Storage Temperature
   -20 °C to +60 °C
Dust/water resistance
   Meets IP52, IEC 529
Operating Altitude
   2800 m
Note
*  Except frequency and AC

 

Safety Specifications
Agency Approvals
   CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993

 

Mechanical & General Specifications
Size
   130 x 236 x 61 mm
Weight
   1.4 kg
Batteries
NiCd:  7.2V, 1.7 Ah
Battery Life
   ~8 hours typical
Battery Replacement
   Via snap-shut door without opening calibrator; no tools required
Side port connections
   Pressure module connector
   Connection for optional battery eliminator
Data storage capacity
   1 day of calibration results
90 day specifications
   The standard specification intervals for the 741B is 1 and 2 years.
   Typical 90 day measurement and source accuracy can be estimated by
    dividing the one year "% of reading" or "%of output" specifications by 2.
   Floor specifications, expressed as "% of full scale" or
    "counts" or "ohms" remain constant.

Permintaan Penawaran